Using 5G NR Signal Analyzer
196
Multi-ACLR measurement with 5G NR
NOTE
You can set the Lowest Ref. Frequency and Highest Ref. Frequency by tapping the rectangle with value using the
on-screen keyboard.
Spurious emissions
The Spurious Emissions measurement is to identify or determine the power level of in-band or out-of-band spurious emissions
within the necessary channel bandwidth and modulated signal. The CellAdvisor 5G indicates either Pass or Fail based on the
specified limit of the signal. Frequency extended up to 18.5 GHz in addition to existing 24 GHz to 40 GHz.
Setting measure setup
After configuring spectrum measurement in "
Configuring Spectrum Measurements
", you can set the measure setup to continue your
measurement. The frequency setting is not used in the Spurious Emissions mode. The measurement settings can be saved and
recalled as a file by selecting File Type as Setup and load the file onto the instrument using the
Save/Load
icons on the side bar.
1
Tap the
Setup
(
)icon on the side bar.
2
Tap
BS Type
and select the option from
1-C/1-H
,
1-O
, or
2-O
.
Set 1-C (Conducted)/1-H (Hybrid) for multi-band operation with mapping of transceivers to one or more antenna connectors (1-
C) or TAB (transceiver array boundary) connectors for 1-H. Set 1-O when BS operates at FR1 and 2-0 when BS operates at
FR2.
3
Tap
Measure Type
and select the option from
Transmitted
or
Receiver
.
4
Tap to switch
Setting
to
User
if you want user defined setting and tap
Configuration
or
Configuration
(
)icon.
a
Tap
Range
under the chart screen and switch to
On
to display or
Off
to hide the selected range in the result table.
You can select the range number between
1
and
20
to add as a new or change the existing settings.
b
Tap
Start Frequency/Stop Frequency
and enter the value for the selected range using the on-screen keyboard.
c
Tap
Start Limit
/
Stop Limit
and enter the lower limit/upper limit for Pass/Fail indication.
d
Tap
Attenuation
/
RBW
/
VBW
and specify or select the value.
5
Tap to switch
Setting
to
3GPP
if you want 3GPP standard defined setting and tap
Configuration
or
Configuration
(
)icon.
a
Tap
Range
under the chart screen and switch to
On
to display or
Off
to hide the selected range in the result table.
You can select the range number between
1
and
20
to add as a new or change the existing settings.
Содержание CellAdvisor 5G
Страница 1: ...1 CellAdvisorTM 5G User s Guide ...
Страница 2: ......
Страница 18: ......
Страница 22: ......
Страница 110: ......
Страница 146: ......
Страница 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Страница 264: ......
Страница 312: ......
Страница 332: ......
Страница 338: ......
Страница 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Страница 352: ......
Страница 356: ...Appendix 344 ...
Страница 357: ...345 ...
Страница 358: ......