Using Spectrum Analyzer
48
Access
:
on the Quick access and display tab or
Menu
>
Limit
.
Limit
The following is the definition of each key function:
Parameter
Definition
Display Line
Tap to set values for the reference line using the on-screen keyboard.
Access
:
Menu
>
Limit
>
Display Line
Display Line
Mode
On/Off
Tap to set the
Display Line Mode
to
On
or
Off
to display or dismiss the reference line. The inactive vertical
move bar on the left edge with straight line appears across the screen to be used as a visual reference only.
To adjust the display line, tap and drag the move bar up or down, then the vertical move bar becomes purple
and activated. If there is no touch operation for 2 seconds, the move bar becomes inactive again. This
display line does not indicate Pass/Fail.
Access
:
Menu
>
Limit
>
Display Line Mode
On/Off
Multi
Segment
Line
Tap
Multi Segment Line
for Pass/Fail indication.
Access
:
Menu
>
Limit
>
Multi Segment Line
Multi segment line
1
Tap
Menu
>
Limit
.
2
Tap
Multi Segment Line
.
3
Tap to switch the select side to
Upper
or
Lower
.
4
Tap the
# of Line
to set the number of segments for the selected side for upper of lower up to 50 segments using the on-screen
keyboard.
5
Tap
Autoset
soft key to automatically set the limit for each segment and display the line.
6
Tap
Offset Up/Down
or
Offset Left/Right
to move the limit line using the on-screen keyboard.
The following table shows the parameters displayed for the multi-segment line.
Parameter
Definition
Autoset
Tap to autoset the limit for each segment and display line.
Access
:
Menu
>
Limit
>
Multi Segment Line
>
Autoset
Select Side
Tap to set the
Select Side
to
Lower
or
Upper
to select the one to be displayed.
Access
:
Menu
>
Limit
>
Multi Segment Line
>
Select Side Lower/Upper
Limit
Tap to select the Limit to
On
or
Off
. Once the
Autoset
is selected, the Limit changes to On.
Access
:
Menu
>
Limit
>
Multi Segment Line
>
Limit
On/Off
# of Line
Tap to set the
# of Line
up to 50 segments for the selected upper or lower limit line using the on-screen
keyboard.
Access
:
Menu
>
Limit
>
Multi Segment Line
>
# of Line
Offset
Up/Down
(dB)
Tap to set the
Offset Up/Down
to move the limit line up or down when the Display Line Mode is set to on
using the on-screen keyboard.
Access
:
Menu
>
Limit
>
Multi Segment Line
>
Offset Up/Down
Offset
Left/Right
Tap to set the
Offset Left/Right
to move the limit line left or right when the Display Line Mode is set to on
using the on-screen keyboard.
Access
:
Menu
>
Limit
>
Multi Segment Line
>
Offset Left/Right
Содержание CellAdvisor 5G
Страница 1: ...1 CellAdvisorTM 5G User s Guide ...
Страница 2: ......
Страница 18: ......
Страница 22: ......
Страница 110: ......
Страница 146: ......
Страница 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Страница 264: ......
Страница 312: ......
Страница 332: ......
Страница 338: ......
Страница 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Страница 352: ......
Страница 356: ...Appendix 344 ...
Страница 357: ...345 ...
Страница 358: ......