Using Spectrum Analyzer
56
c
Select
Integration BW
to set the channel power bandwidth for the selected offset using the on-screen keyboard.
d
Select
Lower
to set the threshold for the power difference between the main channel bandwidth and the selected offset
bandwidth on the left using the on-screen keyboard.
e
Select Upper to set the threshold for the power difference between the main channel bandwidth and the selected offset
bandwidth on the right using the on-screen keyboard.
3
Tap
Main Channel BW
to set the main channel bandwidth and input the value using the on-screen keyboard.
You can also tap the rectangle with value,
under the chart screen and input the value using the on-
screen keyboard.
4
Tap
Menu
>
Sweep
>
Gated Sweep
to set the Gated Sweep
On
or
Off
. This function needs License S015.
5
Tap
Menu
>
Trace
>
Threshold RMS
to switch
On
or
Off
. Threshold RMS function detects RMS above a certain level of power
and make sample detecting for the power below that level. When switched
On
, Average is automatically set to 100 while
switched
Off
, Average is automatically set to 1. Threshold can also be adjusted by the menu under
Trace
,
Threshold Auto
and
Threshold RMS Level
.
An offset represents a difference in center frequencies of the carrier channel and its adjacent channel to be measured. The
frequency range for each offset is specified with an Offset Frequency and Integration BW. Each offset that is created has a Lower
and Upper set of frequencies.
NOTE:
Make sure the Detector is selected to RMS when you perform the adjacent channel power measurement.
Setting limit
1
Tap
Menu
>
Limit
.
2
Tap to switch the
Test Limits
to
On
or
Off
to enable or disable Pass/Fail indication.
The result table color for Pass is green, and the result table color for Fail is red.
3
Optional.
Tap the Save hot key on the side bar and select the save option form the choices available for your measurement
mode.
Adjacent channel power measurement with spectrum analyzer
Multi adjacent channel power
The Multi-ACP measurement is used to do multi-channel ACP testing. It helps you to measure ACP in multichannel transmitting
base station environment.
Setting measure setup
After configuring spectrum measurement in "
Configuring Spectrum Measurements
", you can set the measure setup to continue your
measurement. The measurement settings can be saved and recalled as a file by selecting File Type as Setup and load the file onto
the instrument using the
Save/Load
icons on the side bar.
Содержание CellAdvisor 5G
Страница 1: ...1 CellAdvisorTM 5G User s Guide ...
Страница 2: ......
Страница 18: ......
Страница 22: ......
Страница 110: ......
Страница 146: ......
Страница 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Страница 264: ......
Страница 312: ......
Страница 332: ......
Страница 338: ......
Страница 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Страница 352: ......
Страница 356: ...Appendix 344 ...
Страница 357: ...345 ...
Страница 358: ......