4.8 Menu of general ventilation parameters setting
4-25
4.8.3 Menu of iSV mode parameters
Designation of
parameter
Description of parameter
Values
Vtmax calc.
coeff.
Sets the level of inspiratory volume limitation Vtmax, at that
Vtmax is calculated automaticallyVtmax = Vtmax calc. coeff. x
IBW, where IBW – ideal body weight (see it.4.8.1).
(default value 22 ml per kg)
7 - 30
ml per kg
Enable IRV in
iSV
Enable/disable of the inverse mode of ventilation when inspira-
tion is longer than expiration (default value off).
off/on
4.8.4 Menu of Pmax, Tinsp limits
Designation of
parameter
Description of parameter
Values
Pmax limit
Allows to increase the upper limit of Pmax to 105 cmH
2
O (line
is available only in the CMV/VCV, CMV/PCV, SIMV/VC,
SIMV/PC, SIMV/DC, PCV/VG modes at the adult type of pa-
tient)
87, 105
cmH
2
O
Tinsp limit
Allows to increase the upper limit of inspiratory time Tinsp to 10
s (line is available only in the CMV/VCV, CMV/PCV, SIMV/VC,
SIMV/PC, SIMV/DC, PCV/VG modes at the adult type of pa-
tient)
3, 10 s
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...