4.8 Menu of general ventilation parameters setting
4-23
Designation of
parameter
Description of parameter
Values
After exit from the NIV mode the old value of ETS is
restored. ETS in NIV is measured in lpm.
Sigh
Mode of the periodic deepen sigh.
3
on, off
Trigger Vinsp 25ml
Additional criterion of activation of the inspiratory
trigger, when inspiratory volume exceeds 25 ml. That
provides the triggering at the slow spontaneous
inspirations.
on, off
Sens. to disconn.
Sensitivity threshold of device to the
"
Disconnection
"
alarm. It is defined as the inspiration to expiration
volume
ratio
expressed
in
percent.
Alarm
"
Disconnection
"
will be formed when the inspiration
to expiration volume ratio is below the sensitivity to
disconnection.
0…50 %
(step 5 %)
Plateau time
Selecting of the parameter unit
%, s
Т
disconn.in NIV
Delay time of disconnection alarm triggering in the
NIV mode. Default parameter value is 5 seconds.
0…60 s
1 – If user tries to enter a patient's height <57 cm, the ventilator goes into the patient's weight
editing mode (to the minimum value of 2 kg). The reverse transition occurs when user enters a
weight > 5 kg. At weight values between 2.0 and 5.0 kg, the iSV mode becomes unavailable
2
–
Applied method of calculation of compliance and resistance is described in Appendix 2.1. To
measure compliance immediately, press the compliance window on the device’s display. That is
sensor button at the same time.
3 – Parameters of the periodic deepen sigh mode: periodicity - every 50 respiratory cycles
(inspiration), tidal volume - 1.5 times more than set one (for ventilation modes with volume
control), maximum pressure – on 5 cmH
2
O more than the set one (for ventilation modes with the
pressure control).
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...