Appendix 1.1 Description of CMV / VCV mode
App.1-1
APPENDIX 1 DESCRIPTION OF VENTILATION MODES
Appendix 1.1 Description of CMV / VCV mode
In CMV/VCV mode (Controlled Mandatory Ventilation / Volume Control Ventilation) the patient
receives the set tidal volume. Flow curve is rectangular or decreasing (as shown below), ac-
cording to doctor’s choice. The device provides the set volume of a respiratory cycle at the set
respiratory rate with the set duration of the respiratory cycle.
Figure. App. 1.1 – Graphs of pressure (top) and flow (bottom) in CMV / VCV modes with de-
creasing flow. The shaded area shows volumes of both inspiration and expiration
In the CMV/VCV mode possibility of support of the spontaneous breaths initiated by the patient
is provided. This possibility is realized by setting of the trigger sensitivity. Automatic breath is
triggered when trigger operates within a trigger window.
In CMV/VCV modes user also can set the inspiration pause by Tplat parameter.
Paw
t
PEEP
Flow
t
Respiratory cycle with com-
pliance measurement
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...