Appendix 7 ADDITIONAL PARAMETERS AND CHARACTERISTICS
App.7-3
No
Parameter
Value (description)
3.12.
Endotracheal tube compensation, ETC
0 - 100 %
4.
Respiratory monitoring parameters
4.1.
Mean pressure for the respiratory cycle
Pm
4.2.
Residual pressure level in lungs
AutoPEEP
4.3.
Minute volume of spontaneous breaths
MVspont
4.4.
Frequency of spontaneous breaths
fspont
4.5.
Dynamic compliance/ resistance
C, R (LSF)
4.6.
Peak inspiratory flow
FlowPeak
4.7.
Leakage flow from the breathing circuit
Leak
4.8.
Plateau pressure
Pplat
4.9.
Elimination of
СО
2
per minute (option)
VCO
2
4.10.
Minute alveolar ventilation, alveolar ventilation (option)
MValv, Valv
4.11.
Functional dead space (option)
Vd
4.12.
Cardiac output according to Fick (option)
С
O
4.13.
Auxiliary external pressure (option)
Transpulmonary pressure (option)
Paux
Ptp
5.
Extended respiratory monitoring (RESP2 window)
5.1.
True pressure level in lungs at the end of expiration
(pressure in the lungs by the end of exhalation taking in-
to account the incomplete expiratory PEEtot = PEEP +
AutoPEEP)
PEEPtot
5.2.
Residual pressure level in lungs (that occurs due to the
incompleteness of exhalation)
AutoPEEP
5.3.
Flow at the end of expiration (residual gas flow from the
lungs arising from incomplete exhalation)
ExpEndFlow
5.4.
Expiratory time constant (time constant that determines
the potential rate of pressure change in the lungs during
expiration
τ
exp = Rexp × Cexp)
RCexp
5.5.
Inspiratory time constant (time constant that determines
the potential rate of pressure change in the lungs during
inspiration
τ
insp = Rinsp × Cinsp)
RCinsp
5.6.
Stress index. The coefficient describes the adequacy of
PEEP and Vt selection. Can be defined in the
CMV/VCV and SIMV/VC modes for hardware breaths at
the rectangular waveform.
SI
5.7.
Respiratory effort index. Parameter characterizes the
strength of the patient's breathing attempts. Measured
in cmH
2
O
P0.1
5.8.
Work of the patient breathing. Parameter characterizes
the work of breathing of the patient based on 1 liter of
tidal volume
Wspont
5.9.
Work of the ventilator breathing. Parameter characteriz-
es the work of breathing of the ventilator based on 1 liter
of tidal volume
Wvent
6.
Extended respiratory monitoring (RESP 3 window)
6.1.
Inspiratory time, including spontaneous (time of the last
inspiration, made by ventilator or patient)
Tinsp
6.2.
Factor of breathing cycle filling (ratio of inspiratory time
to total respiratory cycle time)
Tinsp/Ttot
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...