4.4 Starting window. Beginning of ventilation
4-8
4.4.3 Automatic calculation of initial ventilation parameters
According to the entered data in starting window (height and gender of the patient) the device
ventilation automatically calculated starting parameters for all provided modes.
MV for adults and children differs and is calculated in liters per kg of ideal weight (IBW):
•
For the adult patients MV = 0.1 liter per kg IBW,
•
For the children with weight more than 30 kg MV = 0.1 liter per kg IBW,
•
For the children with weight up to 5 kg inclusive MV = 0.3 liter per kg IBW,
•
For the children with weight from 5 to 30 kg MV = (–0.008×IBW+0.34) × IBW.
The above described calculations are visually presented on graphics.
Figure 4.3 – Automatic MV calculation for adults and children
These calculations are real for a resting condition (without physical activity) for conditionally
healthy person. At the increased physical activity, at the increased body temperature, at the
raised metabolism, at the expressed pain syndrome minute ventilation can increase many
times. At the same time in the condition of the lowered metabolism (for example, under anes-
thesia, etc.), minute ventilation can decrease noticeably. Further correction of minute volume
during ventilation is carried out by the doctor, according to the information of gas composition of
blood, EtCO
2
and clinical condition of the patient.
Values of calculated starting ventilation parameters are given below. These parameters are
starting, guaranteeing the safe start of ventilation.
Table 4.2. Values of the calculated starting ventilation parameters for the new patient
Parameter
Value
Notice
MV
0.1 liter per kg IBW
adults and children with weight more
than 30 kg
0.3 liter per kg IBW
(–0.008×IBW+0.34) liter per kg
IBW
children with weight up to 5 kg
children with weight from 5 to 30 kg
Vt
7.5 ml/kg × IBW
standard value of 7.5 ml/kg is
available to adjust in the [Menu]
→
[Ventilation parameters]
→
[Patient’s
parameters] in the range 6 … 9 ml/kg
IBW with the step 0.5 ml
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...