4.17 Alarms
4-64
No
Alarm mes-
sage
Alarm conditions
Reaction of ventilator,
operator’s actions
6 Capnometer off
(Capnometer off)
No connection with the capno-
meter
Turn on the capnometer through the menu.
After restoring of communication message
disappears. If capnometer does not operate,
contact service department. Further opera-
tion of the device is possible without the
mainstream capnometer.
7 Capnometer er-
ror
(Capno error)
The capnometer error occurred. Further operation of the device is possible
without the capnometer. Contact service
department.
8 No CAN-Eth
module (No
CAN-Eth mod)
No connection with CAN-Eth
module
Further operation of the device is possible
without the CAN-Eth. Contact service de-
partment.
9 No SpO
2
module
(No SpO
2
mod-
ule)
No connection with pulse oxi-
metry module
Further operation of the device is possible
without the pulse oximetry sensor. Contact
service department.
10 SpO
2
module not
initialized (SpO
2
not init)
The pulse oximetry module is
not initialized during the device
switching on
The module is missing. Contact service de-
partment.
11 Capnometer not
initialized (Cap-
no not init)
The capnometer is not initia-
lized during the device switch-
ing on.
The module is missing Contact service de-
partment.
12 Calorimeter not
initialized (Ca-
lor.not init)
The metabolism measurement
module is not initialized during
the device switching on.
The module is missing. Contact service de-
partment.
13 CAN-Eth not in-
itialized(CAN-
Eth no ini)
The CAN-Eth is not initialized
during the device switching on.
The module is missing Contact service de-
partment.
14 No Calorimeter
module
(No Calorimeter)
No connection with metabolism
measurement module.
The module is missing Contact service de-
partment.
15 Trends failure
(Trends failure)
The trends access malfunction
occurred.
The memory card error occurred. It is not
possible to view and save trends, does not
affect on the device operation. Contact ser-
vice department.
16 USB: data media
is full (USB: no
space)
The USB data storage is full
Replace the USB Mass Storage with a new
one with enough free space.
17 Update Pa-
CO
2
/PvCO
2
(Update CO da-
ta)
The PvCO
2
and / or PaCO
2
val-
ues were not updated for more
than 4 hours or the EtCO
2
value
changed for more than 30%
from EtCO
2
measured during
the calculation of cardiac output
Check the patient condition. Update the
PvCO
2
and / or PaCO
2
values.
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...