5.1 Mainstream CO2 sensor (capnometer)
5-1
2
5
OPTIONAL FEATURES OF THE DEVICE
5.1 Mainstream CO
2
sensor (capnometer)
External module of breathing gas analysis without sampling (hereinafter – mainstream capno-
meter) is optionally delivered with the device and allows to monitor:
•
СО
2
concentration at the end (Et
СО
2
) and at the beginning (FiCO
2
) of the exhalation;
•
capnogram (PCO
2
).
Capnometer operating principle is based on the method of measuring the infrared light absorb-
tion in the absorption spectrum of carbon dioxide.
Only mainstream CO
2
sensors supplied by Triton Electronic Systems Ltd.
are compatible with ventilator.
The module includes the mainstream CO
2
sensor and ventilation adapters for different types of
patients:
1.
Enclosure of the sensor;
2.
Ventilation adapter (adult and neonate);
3.
Socket for the connection to the ventilator;
4.
Cable of the sensor.
Figure 5.1 - Mainstream capnometer
It is not recommended to connect an mainstram capnometer when the de-
vice is switched on.
Connection of the mainstream capnometer:
•
Check the sensor window and ventilation adapter. They must be dry and clean. Clean or re-
place the adapter if necessary.
•
Connect the adapter with
СО
2
sensor - combine labels on the adapter and sensor and con-
nect till latching (Figure 5.2).
Figure 5.2 – Connection of CO
2
sensor and adapter
•
Connect the CO
2
sensor to the connector
"
CO
2
"
port on the front panel (Figure 5.3).
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...