4.17 Alarms
4-61
No Alarm message
Alarm conditions
Reaction of ventilator,
operator’s actions
20 Accumulator is
discharged
(Acc Discharged)
The accumulator is fully dis-
charged. Accumulator’s charge
is insufficient to continue the
normal operation of the device
Connect the device to the mains power to
charge the accumulator. The alarm message
will be removed after beginning of the charg-
ing. Contact service department in case of
failure.
21 Less than 10 mi-
nutes left
(< 10 min left)
The accumulator is fully dis-
charged. Accumulator’s charge
is insufficient to continue the
normal operation of the device.
Prior to the full discharge is less
than 10 minutes.
Alarm disappears in the accumulator charg-
ing mode (mains power).
Immediately connect the device to the mains
power to charge the accumulator. The alarm
message will be removed after beginning of
the charging. Contact service department in
case of failure.
22 Accumulator
malfunction
(Acc. malfunct.)
The accumulator malfunction
occurred.
When the device is switched on, the operator
will be offered to switch to the technical fail-
ure mode or continue operation with the risk
of a sudden shutdown, see it. 3.4.
23 Low O
2
concen-
tration
(Low
О
2 )
The measured value of the
oxygen concentration FiO
2
is
below set by the operator dev-
iation (1 - 50 %).
Check the connection and the supply of oxy-
gen to the device. If necessary change the
limit of O
2
tolerance. Alarm automatically
disappears when the oxygen concentration
is returned to the normal range.
24 High O
2
concen-
tration
(High
О
2 )
The measured value of the
oxygen concentration exceeds
the set value on the specified
value.
Alarm disappears when the oxygen concen-
tration returns to the normal range.
25 Fan malfunction
(Fan malfunct.)
The electronic unit fan malfunc-
tion occurred.
Contact service department. The device may
shortly continue normal operation. Alarm dis-
appears after restoring of the normal opera-
tion of the fan.
26 Sound malfunc-
tion
(Sound malfunc.)
Faulty dynamic transducer
(breakage).
Further operation of the device is impossible.
Contact service department.
27 FiO
2
sensor mul-
fanction
(FiO
2
s. malf.)
The ventilation controller
FiO
2
sensor malfunction oc-
curred.
The function of Fi
О
2
level monitoring is not
available. There is no change in Fi
О
2
level
control. The device can operate for a short
period of time. Contact service department.
4.17.2 Medium priority alarms
Table 4.16. Medium priority alarms
No Alarm message
Alarm conditions
Reaction of ventilator,
operator’s actions
Physiological alarms
1 High RB
(High RB)
Controlled respiratory rate is
above than the specified top
threshold.
The alarm is automatically disabled when the
respiratory rate is equal or less than alarm
threshold.
If necessary, reduce the respiration rate, or
adjust the alarm threshold. If autotriggering
is suspected, increase the trigger threshold
of the inspiratory trigger.
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...