6.2 Visual inspection and test of controls
6-2
6.2 Visual inspection and test of controls
The device shall not have mechanical damages and signs of liquid ingress. The shell of the
power cord shall not have any damage or sharp bends.
Buttons and encoder shall clearly respond calling the appropriate action. There shall be no false
responses. If the control touch button area you press is offset relative to its graphical designa-
tion calibrate the touch panel through [Menu]
→
[Service menu]
→
[Calibration]
→
[CIVL]
→
[Touch screen calibration].
6.3 Cleaning and replacement of fan filters
Remove the plastic grid of the cooling fan and fresh air fan on the rear panel of the device (Fig-
ure 2.2, positions 2, 8), easily pulling them toward to you. Remove the dust filters and clean
them by washing in soap solution, rinse in clean water and dry. In the case of their substantial
deterioration use the new dust filters supplied (2 pcs.) or order them through the service organi-
zation.
Make sure that filters are completely dry, set them back in their place and snap the plastic grids
of the fans.
Figure 6.1 – Cleaning and replacing of the dust filter
6.4 Check of sealing gaskets
Check the condition of the sealing gaskets of the:
- oxygen hose;
- water trap fittings of sidestream capnometer (indirect calorimeter).
The sealing gaskets should not have any visible damage or undue strain.
If necessary replace the gasket ordering them at the manufacturer.
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...