4.20 Nebulizer
4-73
It is possible to turn the nebulizer on only if the device operates in one of the ventilation modes.
If you turn the nebulizer immediately at the power supply, nothing will happen.
During inhalation in the indication area gray information message "Nebulizer is on" is displayed.
After the set time interval inhalation stops automatically, and the message disappear.
For early termination of the inhalation select "Nebul. work time 0 min" in [Menu]
→
[Additional
functions].
When user turns the nebulizer on again, countdown will start working.
The device can be provided with blocking of nebulizer in pediatric mode. In addition operation of
the nebulizer is blocked in ventilation mode with the volume control with set V
T
of less than 100
ml.
For the operation in the pediatric mode, the nebulizer with
vibrating mesh
without introducing
additional flow is recommended (AeroNeb PRO), which can be supplied on request in the deli-
very kit of the device. Such nebulizer does not require synchronization with the device and op-
erates independently. All controls are located on the nebulizer.
Figure 4.15
–
Connecting of the nebulizer AeroNeb Pro
During the operation in every breathing cycle the device delivers pressure of 1.8 bar into the
spray chamber of nebulizer. Pure oxygen is used to create the pressure.
At every inspiration medicines are sprayed from the inhaler container. Since spraying is carried
out with oxygen, FiO
2
value of the inspired gas mixture slightly increases. Therefore during inha-
lation, it is necessary to consider an increase in oxygen concentration in air mixture.
Due to its principle of operation the nebulizer creates additional inspiratory flow. In the modes
with volume control the device software takes into account this additional flow.
The software of the ventilator blocks the nebulizer if set tidal volume is less than the additional
flow generated by the nebulizer. In this case it is necessary to transfer the ventilator to operation
with the pressure control.
In the absence of pressure in oxygen line and if the nebulizer is turned on, the display will show
that the nebulizer is on, however, compressed oxygen will not enter the chamber because of its
absence. So spraying of the medicines by the nebulizer will not occur.
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...