4.14 Alarm settings menu
4-45
Line of the
menu
Designation, description of parameters
Possible values Default value
EtCO
2
_max
– top alarm threshold of
СО
2
concentration in exhaled air [%].
EtCO
2
_min…12
%
6
EtCO
2
_min
– bottom alarm threshold of
СО
2
concentration in exhaled air [%].
2… EtCO
2
_max
%
2
Change measuring units
– switches unit of
measurement of
СО
2
%, mmHg
mmHg
Pulse
oximeter
SpO
2
_max
– top alarm threshold of SpO
2
–
hemoglobin saturation of arterial blood by oxygen
(hyperoxia)
from 90 % or from
SpO
2
_min
to
100 %
100
SpO
2
_min
– bottom alarm threshold of SpO
2
-
hemoglobin saturation of arterial blood by oxygen
(hypoxia)
60... 95 %
(SpO
2
_max)
86
PR_max
– top alarm threshold of pulse rate
(taxycardia)
80 (PR_min)…
350
1
/min
90
PR_min
– bottom alarm threshold of pulse rate
(bradycardia)
15…100
(PR_max)
1
/min
50
Low pulse volume alarm
on/off
off
Changing and saving of parameters is carried out by the encoder. Values of alarm thresholds
remain in non-volatile memory and are restored at power up in previous patient mode, except
alarms "
FiO
2
sensor
", "
FiO
2
alarms
", "
О
2
pressure (min)
".
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...