5.5 Metabolism measurement module
5-16
* -
FiO2 readings may appear with a delay of a few seconds.
•
During using the metabolism measurement module, it is recommended to display PCO
2
, PO
2
[%] window, where capnogram and oxigram of the respiratory cycle (FiO2(t) - EtO2(t)) are
displayed simultaneously.
Figure 5.14 - Combined graph of capnogram and oxigram
•
At displaying of metabolic parameters VCO
2
, VO
2
, RQ, REE averaging is performed in the in-
terval of 1 min.
•
VO
2
, VCO
2
, RQ, REE parameters are stored in the trend and can be viewed later (see it.
4.13).
Metabolism status messages
Metabolism status messages are displayed only in the bottom line of the
“Metabolism” window. These messages are not displayed neither in the
area of alarm list indication nor ‘Alarm log’ section.
Table 5.1. Metabolism status messages
No Alarm message
Alarm conditions
Reaction of ventilator,
operator’s actions
1. Measuring
Measurement
Metabolism module is working properly.
2. Zero calibr.
Zero calibration
The message disappeared automatically af-
ter calibration finishing.
3. Warming
Warming
The message disappeared automatically af-
ter warming finishing.
4. Occlusion
Occlusion
Check the sampling line and water trap for
occlusion. If it is necessary replace the
sampling line with a new one.
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...