5.2 Volume capnometry
5-5
5.2 Volume capnometry
Optionally the device may be equipped with the function of volume capnometry, which operates
only together with mainstream capnometer. Volume capnometer (sometimes the “volumetric
capnometer” term is used) allows to define the following parameters:
•
elimination of CO
2
per minute (VCO
2
), which characterizes the metabolism rate (e.g., high
value indicates sepsis etc.), and reflects the effectiveness of applied respiratory therapy in
dynamics;
•
functional dead space (Vd);
•
minute alveolar ventilation (MValv).
The calculation method of the CO
2
amount per expiration is the combined integration of patient
flow curve and the capnogram from the mainstream capnometer (Figure 5.5).
Figure 5.5 – Synchronization of the capnogram with the flow curve
СО
2
elimination per breathing cycle is:
∫
∫
⋅
⋅
−
⋅
⋅
=
2
1
pat
3
2
pat
)
(
2
)
(
)
(
2
)
(
2
t
t
t
t
i
dt
t
CO
t
F
dt
t
CO
t
F
VCO
(1)
where:
VCO2i –
СО
2
elimination per i-breathing cycle, units, ml,
СО
2(t) – the relative concentration of CO
2
,
F
pat
(t) – the patient's flow rate, lpm.
Measuring units of VCO
2
are milliliters given to the patient’s condition (BTPS). Such reduction is
necessary for the correct application of the formulas for calculating other parameters.
Elimination of
СО
2
per minute is:
∑
=
=
n
i
m
i
VCO
VCO
1
2
2
(2)
where:
VCO
2
m – CO
2
elimination per minute,
n – number of breaths during the last minute.
t1
t2
t3
Flow, Fpat(t)
Capnogram, CO
2
(t)
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...