5.5 Metabolism measurement module
5-15
•
Join the airway adapter with mainstream capnometer and connect it to the patient's tee in the
breathing circuit.
•
Connect the monitoring line adapter to the airway adapter of the mainstream capnometer.
•
Connect the monitoring line to the adapter on the breathing circuit and to the water trap on
the ventilator*.
* - use the monitoring line from the delivery set or similar (internal diameter 1.2 mm, length 1.8 ... 2 m
with two connectors of the Luer-Lock type at the end).
During preparing for the operation observe the following guidelines:
Make sure that all elements of measurement scheme are securely con-
nected to each other. The measurement results are not valid If the lea-
kage occurred.
•
Use only clean and dry sampling lines, without drops of liquid inside.
•
For dripping of the condensate produced in connecting device middle part of it should be
hung higher that its ends. Port of the adapter connected to the sampling line should be faced
up to eliminate the accumulation of moisture in the adapter (see Figure 5.13).
Figure 5.13 – Connecting of the sampling line
Turning on the metabolism measurement module
•
Before switching on the metabolism measurement module connect and warm up the main-
stream capnometer (see it.5.1).
•
Select row “Metabolism” in the “Monitoring” window to turn on the metabolism measurement
module (Figure 4.4, position 14):
•
Check the appearance of the metabolic parameters display window and the status message
"Measurement" in the bottom line of the window:
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...