4.1 Main features of device
4-2
The built-in accumulator allows the device to operate without interruption (at the total charge)
not less than 4 hours in case of absence of the centralized power supply at any ventilations pa-
rameters. At the minute ventilation volume of 10 lpm the device operates from the accumulator
not less than 6 hours.
The transport cart (mobile trolley) allows to move quickly the device to the necessary place, and
also to provide ventilation of the patient during intrahospital transportation.
The device can be used also without the transport cart.
The device operates with the open circuit - gas exhaled by the patient is not used repeatedly.
At installation in the breathing circuit up to 3 virus-bacterial filters pressure in the patient tee
changes no more than on 3 cmH
2
O, at the flow 60 lpm in case of timely replacement of filters at
their remoistening.
The device provides possibility of independent breath of the patient if normal ventilation is im-
possible because of troubles with electro - and pneumo supply. Pressure drop in the inspiration
and exhalation lines, measured in the patient connection port does not exceed 6 cmH
2
O.
The device can operate with oxygen sources of high and low pressure, for example with oxygen
concentrators. The device supports the declared accuracy of FiO
2
only if operates from high
pressure oxygen sources. At operation from low pressure oxygen source, accuracy of FiO
2
is
defined by the low pressure oxygen source.
Содержание MV200
Страница 2: ......
Страница 8: ......
Страница 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Страница 124: ...4 21 Open valve function 4 76 For your notes ...
Страница 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Страница 160: ...7 Troubleshooting 7 2 For your notes ...
Страница 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Страница 166: ...13 commission date mark 12 2 For your notes ...
Страница 168: ...14 Maintenance and repair data 14 2 For your notes ...
Страница 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Страница 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Страница 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Страница 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...