5
•
Installation
Page 74
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
NOTICE
Be aware of the backpressure limit of the flow cell in the Vanquish UV/
VIS detector connected in the flow path before the fluorescence
detector. Connect the capillary from the UV/VIS detector directly to the
fluorescence detector flow cell inlet. Avoid connecting any additional
components in the flow path between the two detectors. Refer to the
Operating Manual for the UV/VIS detector
.
Figure 24: Connecting the inlet capillary from the diode array detector
(example)
1. Connect the inlet capillary from the column compartment to the
flow cell inlet of the first detector. Refer to the instructions in the
Operating Manual
for the first detector in the Vanquish system flow
path.
2. Route the connecting capillary from the flow cell outlet of the first
detector upward to the fluorescence detector.
3. On the leak tray of the fluorescence detector, pull the capillary clip
carefully to the right to open the clip.
4. Position the capillary behind the capillary clip.
5. Carefully release the clip and take care not to clamp the capillary.
The capillary must be secured behind the clip.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...