Maintenance and Service
Page 132
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
7.5
Performing a Wavelength Calibration
A simple initialization procedure is performed after power-up of the
detector. This requires that a flow cell is installed in the detector. For
details, see
To ensure optimum performance and wavelength accuracy, perform a
wavelength calibration using water (Raman measurement). During
wavelength calibration, the light spectrum of the xenon flash lamp is
used to calibrate the excitation wavelength. Afterward, the emission
monochromator is adjusted with the help of the maximum of the Raman
emission spectrum (397 nm).
When
•
After the flow cell has been installed or replaced
•
After the lamp has been replaced (performed by service engineer)
•
If wavelength validation fails
Preparations
The following conditions
must
be met for both wavelength validation
and Raman wavelength calibration:
•
The system is equilibrated and environmental conditions are stable.
•
There are no air bubbles in the flow cell. We recommend degassing
the water. Set a flow rate of 0.5 mL/min and wait about 15 minutes
before you begin, until the baseline is stable.
•
The water has no impurities. Always use fluorescence-grade water
and consider using a filter or a suitable column before the detector.
To perform a wavelength calibration
You can perform wavelength calibration via Chromeleon. Wavelength
calibration can take a few minutes. During this time, data acquisition is
not possible.
1. Execute the
CalibrateRaman
command to start the calibration.
The following message appears in Chromeleon:
Make sure that
¨
the system is equilibrated and conditions are stable
¨
fluorescence-grade water is flowing through the cell
¨
the water is degassed (no air bubbles) and free of particles
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...