Installation
Page 62
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Parts required
Flow cell
Preparations
1. Remove the cover from the flow cell opening. To do so, loosen the
two finger-tight screws. The screws are captive in the cover and do
not need to be removed.
TIP
Keep the cover to close the flow cell opening when no flow cell is
installed in the device, especially when the detector is transported or
shipped.
2. Unpack the flow cell.
Figure 14: Flow cell of the fluorescence detector
No.
Description
1
Flow cell screws - Used to mount the flow cell to the detector.
2
Flow cell label
3
Outlet
-
Used to connect the waste line.
4
Inlet
-
Used to connect the inlet capillary.
5
5a, 5b, 5c
Optical block - Do not touch the optical block.
Optical ports
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...