6
•
Operation
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 121
Situation
Steps
Device shall be transported or shipped
after shutdown
If one of the modules shall be removed
from the system stack, turn off
all
system modules with their main power
switch. Pressing the system power
button will not be sufficient to turn off
the power to the devices completely.
Follow the instructions in
.
TIP
To prevent dust particles from causing damage to the detector optics
during periods of detector inactivity, install a flow cell or reinstall the
flow cell cover.
See also
2
Connecting the Inlet Capillary (
2
Connecting the Detector Waste Line (
6.10.3 Restart after Long-Term Shutdown
To restart the device after a long-term shutdown, follow these steps:
1. Prepare and restart the other modules in the Vanquish system,
following the instructions in the
Operating Manuals
for the modules.
Pay special attention to the
Preparing the Module for Operation
section.
2. Flush the components in the flow path before the flow cell before
you connect the flow cell to the system flow path.
3. Connect the inlet capillary to the flow cell inlet and the waste line to
the flow cell outlet.
4. Turn on the device. Observe the following:
Situation
Action
If the device remained in the system
stack and all system modules were
turned off.
Turn on the system with the system
power button on the system base.
If the device is restarted after transport.
Turn on the device with the main power
switch.
5. Before starting an analysis, let the detector equilibrate and be sure
that it is ready for operation.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...