6.9.10 Baseline Behavior
If wavelengths are switched or the sensitivity, filter wheel, or PMT
setting are changed during the method, the chromatogram may show
baseline jumps because of the background fluorescence at the new
wavelength setting. Three baseline behavior mode settings define how
the baseline behaves in these cases:
•
Zero
: sets the baseline to zero
•
Append
: appends the baseline to the previous signal
•
Free
: lets the baseline "jump" to the current absolute value
TIP
You can perform
Autozero
command (automatic null balancing) any
time, that is, the intensity measured at the time of the autozero is
subtracted from any intensities measured thereafter. Always select a
time for null balancing at which no sample flows through the flow cell.
The values are stored and can be reset using the
ClearAutozero
command.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...