7
•
Maintenance and Service
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 141
1. Use the screwdriver to remove the fuse holder.
2. Replace the two fuses with new fuses of the specified type and
current rating. Always replace
both
fuses.
3. Reinstall the fuse holder.
4. Reconnect the power cord to the power source and to the device.
5. Turn on the device with the main power switch.