1. Calculate p6 = p5 + p3 + p0.
p6: Pressure at the inlet port of the flow cell
p5: Pressure drop at the flow cell
p3: Pressure drop at the transfer capillary and the additional module
p0: Pressure drop at the waste line
2. Compare p6 with the pressure limit of the flow cell.
Situation
Steps
p6 is at least 5 bar below
the pressure limit of the
flow cell
The configuration can be used.
p6 is less than 5 bar
below the pressure limit
of the flow cell or above
the limit
Operate the system configuration at a reduced flow
rate.
–or–
Change the system configuration (for example, the
transfer capillary) to meet the specification of the flow
cell.
See also
2
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...