6
•
Operation
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 109
•
If the selected sensitivity is
too large
, the PMT signal is saturated. In
this case, the detector automatically reduces the sensitivity. Markers
appear in the chromatogram at the beginning and at the end of the
saturation (see figure,
Sensitivity = 8
). If the detector is not saturated
any more, the Sensitivity setting before saturation is restored.
TIP
When the sensitivity was automatically reduced or when the initial
value was restored, the Chromeleon Audit Trail shows a warning with
the new sensitivity value after the adjustment.
Figure 33: Effects of the Sensitivity setting
With Vanquish fluorescence detectors, the intensity of the emission light
(measured by the PMT) is normalized with the intensity of excitation
light through the sample (measured by the reference sensor). Therefore,
the values (counts) that the emission channel displays cannot be used
for optimizing the sensitivity. In the figure, the PMT is saturated at
63 million counts. However, under different measuring conditions, the
PMT may be saturated at a value of 100 million counts, for example.
Therefore, use the
MaxPMTSaturation
parameter to optimize the
sensitivity for each peak as described below. This parameter
continuously records the maximum PMT saturation since the last
ClearMaxPMTSaturation
command. The result is reported in percent of
the maximum permitted PMT saturation.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...