![ThermoFisher Scientific Vanquish VC-D50 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/thermofisher-scientific/vanquish-vc-d50/vanquish-vc-d50_operating-manual_1105630025.webp)
2
•
Safety
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 25
WARNING—Hazardous Vapors
Mobile phases and samples might contain volatile or flammable
solvents. The handling of these substances can pose health and safety
risks.
•
Avoid accumulation of these substances. Make sure that the
installation site is well ventilated.
•
Avoid open flames and sparks.
•
Do not operate the device in the presence of flammable gases or
fumes.
CAUTION—Escape of Hazardous Substances from PEEK Capillaries
In the Vanquish system, capillaries made of PEEK may be used. Swelling
or attack by acids can cause PEEK capillaries to start leaking or to burst.
Certain chemicals, for example, trichlormethane (CHCl
3
), dimethyl
sulfoxide (DMSO), or tetrahydrofuran (THF) can cause PEEK to swell.
Concentrated acids, such as sulfuric acid and nitric acid, or a mixture of
hexane, ethyl acetate, and methanol, can attack PEEK.
•
Swelling or attack is not a problem with brief flushing procedures.
•
For more information, refer to the technical literature on the
chemical resistance of PEEK.
CAUTION—Allergic Reaction
Some capillaries in the Vanquish system are made of MP35N™, a nickel/
cobalt-based alloy. Individuals with sensitivity to nickel/cobalt may show
an allergic reaction from skin contact.
CAUTION—Sparking due to Electrostatic Discharge
Liquid flowing through capillaries can generate static electricity. This
effect is particularly present with insulating capillaries and non-
conductive solvents (for example, pure acetonitrile). Discharge of
electrostatic energy might lead to sparking, which could constitute a fire
hazard.
Prevent the generation of static electricity near the chromatography
system.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...