Operation
Page 100
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Setting the Multi-Channel Performance
In multi-channel mode, you can select a measuring performance (multi-
channel performance) to determine whether the measurement is
performed faster, but with more noise, or whether minimum noise is
required, which means a longer measuring time and lower data
collection rate. The possible values are:
UltraFast
,
Fast
,
Standard
,
LowNoise
,
UltraLowNoise
.
Response time and data collection rate for multi-channel mode are
determined automatically on the basis of the selected multi channel
performance and the selected parameters, and adjusted each time
wavelengths are switched during the measurement.
Using multiple channels in the Chromeleon software
Use the Instrument Method Wizard (Advanced mode) to define the
parameter settings for all channels and the multi-channel performance.
See also
2
Determining the Optimum Excitation and Emission Wavelength
(
6.7.3
Zero Order Mode
In Zero Order Mode, the grating of the emission monochromator reflects
the entire emission spectrum of the sample onto the PMT, rather than
only a single wavelength. The excitation monochromator is set to a
single wavelength as usual.
Use this mode during method development, if you do not know the
retention times and emission wavelengths of the various substances in
your sample. As the entire range of emission wavelengths is acquired,
you can determine the retention times of all substances in a single run,
as long as the substances can be excited at the selected excitation
wavelength.
TIP
Use the filter wheel setting (VF detectors only) to cut off undesired
wavelengths. For example, set the filter wheel setting to 370 nm to
restrict the emission wavelength range to wavelengths above the filter
wavelength.
The Zero Order Mode is also suitable for samples that emit light in an
exceptionally broad band. In this case, the measured intensity is higher
in Zero Order Mode than under normal operation where most of the
emitted light is discarded by the emission monochromator. This might
allow you to achieve a better limit of detection.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...