Page 107
•
Observe the general guidelines and recommendations on the use of
solvents and additives in the chromatography system. Refer to
Use
of Solvents and Additives
in the
Vanquish System Operating Manual
.
See also
2
Overview of Optimization Parameters (
2
6.9.2
Overview of Optimization Parameters
The following table serves as an overview of parameters that influence
detector performance and indicates the performance characteristics
affected. The sections below offer guidelines for selecting the
parameters.
Parameter
Affects
Excitation wavelength,
emission wavelength
Fluorescence intensity, limit of detection,
selectivity
Sensitivity
Baseline noise, max. fluorescence intensity
Filter wheel setting
Baseline noise, permitted wavelength range for
emission
PMT
Emission wavelength range (and therefore
excitation wavelength range)
Response time
Baseline noise, peak width, peak height
Data collection rate
Peak resolution, disk space, possibly baseline noise
Flow cell temperature
Fluorescence intensity, reproducibility
Lamp mode
Lamp lifetime, baseline noise, maximum data
collection rate
Baseline behavior
Course of the baseline after switching the
wavelength, sensitivity, filter wheel, or PMT.
6.9.3
Determining the Optimum Excitation and Emission Wavelength
The most important parameters that need to be optimized are the
excitation and emission wavelength. Note the following key criteria for
determining the wavelength for an analysis:
•
Preferably, select an excitation wavelength on the absorption
maximum of the sample components.
•
Avoid the wavelength range where the solvents absorb (for
example, below 220 nm for methanol and below 210 nm for
acetonitrile). The excitation wavelength should always be selected
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...