7
•
Maintenance and Service
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 131
7.4.4
Monitoring the Lamp Age
The Chromeleon software supports functions for monitoring the lamp
age. This function can help to decide when a lamp is due to be replaced.
The flash lamp can produce about 1-1.5x10
9
flashes before it needs
replacement only if the performance of the detector is no longer
sufficient. Thus, the lifetime for the flash lamp depends on the flash
frequency, and therefore the selected lamp mode. Based on the number
of lamp flashes, the Chromeleon software calculates a value that
indicates how many hours the lamp was operated. The value is based on
the assumption that the lamp was operated in
Standard
mode. For the
calculation of the expected lamp lifetime, one operating hour is
weighted depending on the selected lamp mode.
Mode
1 operating hour is counted as ...
Expected lamp lifetime
LongLife
0.25 hours
approx. 16000 hours
Standard
1 hour
approx. 4000 hours
HighPower
3 hours
approx. 1300 hours
Example:
If the lamp age in Chromeleon indicates 2000 hours, this
means that the lamp has reached about the half of its lifetime. Thus, if
the lamp is operated in
LongLife
mode, it is expected to operate for
another 8000 hours.
In Chromeleon, check the
EffectiveLampAge
parameter. If the value
exceeds 4000 hours, the lamp should be replaced. Note that the lamp
must be replaced only by a Thermo Fisher Scientific service engineer.
The service engineer will reset the lamp age counter to zero after the
lamp was replaced.
TIP
You can extend the lifetime of the xenon flash lamp by switching
between lamp modes during a chromatographic run.
See also
2
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...