2
•
Safety
Page 28
Integral Fraction Collector FT (VF-F20-A) Operating Manual
2.4
Solvent and Additive Information
2.4.1
General Compatibility
To protect optimal functionality of the Vanquish system, observe these
recommendations on the use of solvents and additives:
•
The system must be used with reversed-phase (RP) compatible
solvents and additives only.
•
Use only solvents and additives that are compatible with all parts in
the flow path.
2.4.2
Allowed pH Ranges
Allowed pH ranges (standard system configuration):
System
(Standard
Configuration)
Allowed
pH ranges
Remarks
Vanquish Core
Limited to
2-12 by
Integral
Fraction
Collector
•
pH
values
of 2 or less
: The application time should
be as short as possible. Flush the system
thoroughly after these applications.
•
pH values higher than 9.5 with optical detectors
:
Avoid using mobile phases with a pH value higher
than 9.5 together with optical detectors. This can
impair the functionality and optical performance of
the detector flow cell.
•
pH values higher than 12
: May affect
electrochemical detection. Before using highly
alkaline solvents for flushing the system,
disconnect the detector from the system.
•
Mobile phases containing ammonium hydroxide
: In
rare cases, a shortened lifetime of reversed-phase
(UHMW-PE) piston seals has been observed with
high pH, ammonium hydroxide containing mobile
phases and prolonged exposure.
Vanquish
Horizon
Vanquish Flex
2-12
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...