Installation
Page 84
Integral Fraction Collector FT (VF-F20-A) Operating Manual
3. Tighten the reservoir cap hand-tight. Press the retaining guide into
the hole in the reservoir cap to ensure that the tubing is kept in
place in the cap.
4. Close any open holes in the reservoir cap with cap plugs.
5. Place the needle wash reservoir in the solvent rack. Position the
needle wash line straight in the tubing guides.
6. Check the needle wash line across the entire flow path:
a) Make sure the line is not bent, pinched or squeezed at any point
in the flow path.
b) If you have to cut tubing to length, use a tubing cutter. Make
sure that the cut is at right angle to the length of the line.
Purging the Needle Wash System
After the device has been turned on, purge the needle wash system
using Chromeleon to fill the needle wash port with the fresh needle
wash liquid. During purging, the needle wash port is flushed
continuously until the fresh needle wash liquid is present.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...