8
•
Troubleshooting
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 207
Message and Code
Description and Remedial Action
Code 9007. Needle vertical drive
homing time out
Needle drive
can’t return home position
.
Check if the needle drive is physically blocked.
After resolving physical blockage, then perform
Self Test
to recover the
issue.
If the issue persists, contact Technical Support.
Code 9008. Carousel blocked. Check for
obstructions with racks or well plates.
Rotate the carousel and check if it can be moved easily. If obstacles are
present in the way of carousel, remove the obstacles.
After resolving physical blockage, then run
Self Test
to recover the
issue.
If the issue persists, contact Technical Support. Check if the carousel is
physically blocked. After resolving physical blockage, then run self-test
to recover the issue.
If error persists, contact Thermo Fisher Technical Support for
assistance.
Code 9009. Vertical needle drive
blocked.
Check if the needle drive vertical needle drive is physically blocked in
vertical direction.
After resolving physical blockage, then run self-test to recover the
issue.
If the issue persists, contact Technical Support.
Code 9010. Vial not found. Check for
correct vial position in the rack and
sample sequence.
Vial not found with
InVial
collection mode.
Check if vials or well plate are present at the specified positions.
Check the Frac Start position (start collection position) in FC ePanel tab,
and then check the settings in the sequence view where individual
collection position is specified.
Specify a different tube position, or place vials to specified positions as
the setting in the ePanel FC tab or sequence view, if necessary.
Check if a sample rack or well plate is present at the specified position.
Place a sample container at the specified position. After resolving
physical blockage, then run self-test to recover the issue.
If the issue persists, contact Technical Support.
Code 9011. Carousel hit an unexpected
object.
Carousel hit an expected object due to improper needle height setting
or improper well plate height, or other physical blockage in the way of
carousel. Carousel stepper motor step loss detected. Check any
unexpected object.
Check if the correct sample rack or well plate is configured and check if
needle height or well plate height is properly set manually. Then run
self-test to recover the issue. Check if the correct sample rack or well
plate is configured
After resolving physical blockage, then run self-test to recover the
issue.
If the issue persists, contact Technical Support.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...