Integral Fraction Collector FT (VF-F20-A) Operating Manual
Parameters
Description
Delay volume
The Fraction Delay Volume is the fluidic
volume travelled by a peak after triggering
the set start threshold conditions between a
detector and the separation point of the
downstream fraction collection valve.
The volume can be logged automatically if the
Vanquish Fraction Collector Delay Volume
Determination
operational qualification is
performed. Before running the OQ test, it is
recommended to perform first an Instrument
Installation Qualification to ensure that the
standard HPLC system is basically ready to
operate in its intended use.
(DVD) (
page 120).
The
Delay Volume
can be changed by
specifying delay Capillary inner diameter (ID)
and length (L).
Delay capillary
IDxL
Select either
Custom
or predefined
Delay
Capillary IDxL
. If the predefined
Delay
Capillary IDxL
is selected
,
the
Delay Volume
is automatically updated.
Delay capillary
ID
Delay capillary L
If
Custom
as
Delay Tubing Capillary IDxL
is
selected, then need to manually input
Delay
Capillary ID
and
Delay Capillary L
. The
Delay
Volume
is calculated and displayed
accordingly.
Fraction collection options
Off (do not
collect at all)
Select this option to disable fraction
collection.
Collect by peak
When this option is selected, fraction
collection starts depending on the specified
detection channels and peak detection
settings.
Collect by time
Fraction collection is independent of peak
detection.
When this option is selected, use the
Collection period
input field to specify the
time in [s] after which a new tube will be
used.
Tube wrapping
Select this check box to return to the first
tube after the last tube is filled. If this check
box is cleared, the Chromeleon Queue will be
aborted when the last tube is reached.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...