6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 113
6.6
Important Operating Parameters
The parameters described in this section should be considered for
routine operation of the device. You can usually access these
parameters from the Chromeleon user interface.
If a parameter listed below is not available in Chromeleon, consider
updating the firmware and Chromeleon version.
For more information, refer to
Chromeleon Help and documents
.
Parameters
Description
Rack type (blue, green, red, yellow)
The rack type specifies the type of sample
rack or well plate that is installed in the
respective segment (
RackType_Blue
, _
Green
,
_
Red
or _
Yellow
).
For details on the rack type settings, see
page 111)
.
Change rack
Rotate the required segment of the carousel
to the front to install or remove a sample rack
or well plate.
Tube position
The tube position identifies the position into
which the next fraction is collected.
The
Tube Position
consists of the color code
for the segment and the position on the
sample rack or well plate, which are
separated by a colon (for example,
B:E8
for
segment B, sample position E8).
Reset start
position
Set the start position to the first available
position of the carousel. The segments are
checked in Red, Yellow, Blue, and Green
order to determine the start position. For
example,
R:A1
should be the new position if
there is a rack or well plate on Red segment.
Temperature control
Enable or disable the temperature control for
the sample compartment. As a standard, the
temperature control is disabled.
Temperature
nominal
The temperature setting defines the setpoint
to which the fraction collector compartment
is to be cooled or heated. The default
temperature is set to 25 °C (
Temperature
Nominal = 25
).
For details, see
page 107)
.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...