8
•
Troubleshooting
Page 210
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Message and Code
Description and Remedial Action
Code 9047. Temperature sensor open -
inside heat sink.
Power cycle the device. If the issue persists, contact Technical Support.
Code 9048. Temperature sensor short -
inside heat sink.
Power cycle the device. If the issue persists, contact Technical Support.
Code 9049. Delay volume
determination failed. Flow cell not
supported by this function.
The flow cell of the detector in the system is not supported by the
delay volume determination function.
See
Vanquish Detector Flow Cells (
page 126)
for supported Vanquish
detector flow cells. Check the detector flow cell and make sure it is in
the support list.
Code 9052. Fraction Valve homing
error.
The fraction collection valve fails to return to home position due to a
temporary communication issue or the valve is worn seriously after
running for a long time.
First, run
Self Test
to recover the issue and check the audit trail for
more information. If not resolved, power cycle the device. If the issue
persists, contact Technical Support.
Code 9053. Wrong vial position.
Invalid
tube position
is provided when setting in ePanel FC tab:
Invalid
tube position
is provided when manually moving needle to one
specific position by selecting
Move Needle Position
button on ePanel
FC tab.
Check the racks and make sure the
Tube Position
and
Move Needle
Position
are in correct format and in range.
Code 9054. Air bubble sensor abnormal
during power up check.
Power cycle the device. If the issue persists, contact Technical Support.
Code 9055. Internal needle rinse failed.
Pump speed is 0.
Flow rate of upstream Vanquish pump is 0.
This may occur when performing
Internal Needle Rinse
via ePanel FC
tab or the flush function during a sequence running.
Turn the Vanquish pump on and set proper flow rate, then redo the
manual internal needle rinse or injection.
Code 9056. Flush number of times limit
reached. System does not flush any
more.
Be aware that the flush pump internal volume is 1000uL and limited
flush times are allowed.
When installing needle capillary (P/N: 6706.1010, ID x L: 0.18 x 415
mm) with flush buffer loop (50 µL), the flush number is 20 times in one
injection.
When installing needle capillary (P/N: 6706.1020, ID x L: 0.25 x 415
mm) with flush buffer loop (100µL), the flush number is 10 times in one
injection.
Code 9057. Needle drive pusher status
has conflict with current method
setting.
Current physical pusher status does not align with the configured
Needle movement mode (InVial, AboveVial). Need to manually adjust
the pusher status to be align with the needle movement mode.
Otherwise the sequence cannot be started.
Code 9058. Door open interrupted BCR
scan
The opening of left door will stop BCR scanning. After the door is
closed, the scanning will restart automatically.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...