5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 75
Figure 25: Flow path connection in a two stacks configuration
For a two stacks configuration with a
Vanquish Horizon or Flex UHPLC
system
the flow path connection are as follows:
No.
Description
Shipped with
1
0.10 x 350 mm, MP35N
System Base Vanquish Horizon/Flex
2
Active or passive pre-heater, MP35N
System Base Vanquish Horizon/Flex
3
Depending on flow rate:
•
Up to 2 mL/min: 0.1 x 550 mm, MP35N
•
2 - 10 mL/min: 0.18 x 550 mm, MP35N
Integral Fraction Collector FT
4
Depending on fraction collection mode
and flow rate
Delay capillary for time-based
fractionation:
•
Up to 2 mL/min: 0.1 x 350 mm, MP35N
•
2 - 10 mL/min: 0.18 x 350 mm, MP35N
Integral Fraction Collector FT
Delay capillaries for peak-based
fractionation
•
Up to 0.5 mL/min: 0.18 x 1200 mm,
MP35N
•
0.5
–
1 mL/min: 0.25 x 1500 mm, MP35N
•
1 - 2 mL/min: 0.5 x 800 mm, PEEK
•
2 - 10 mL/min: 1 x 1000 mm, PEEK
Optional, needs be ordered separately
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...