Maintenance and Service
Page 178
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Figure 78: Removing the needle unit
No.
Description
No.
Description
1
Attachment screw for the
needle unit
3
Vial pusher cover
2
Lock pin
4
Vial pusher
NOTICE
Pay attention to the removing and installation process. Make
sure no extra force or damage to the yellow cable nearby the needle
unit.
1. Unpack the replacement needle unit and remove the protection
parts, including a short tubing on the needle tip and a plug.
2. Position the needle unit on the installation plate. Align the two pins
on the rear side of the needle unit into
the holes on the installation
plate.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...