9
•
Specifications
Page 218
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Type
Specification
Temperature Accuracy
- 2 °C / + 4 °C
Temperature Stability
± 1 °C
Cooling performance
Cooling duration < 20 min when reducing cabin temperature by 10 °C from
ambient temperature 25 °C
Needle Move Time
≤
0.75 s vial to vial with needle above sample container (with 54 position
rack)
≤
1.25 s vial to vial with needle in sample container (with 54 position rack)
HPLC Stack Configuration
In one stack or separated in two stacks
Detectors Supporting
Automated Delay Volume
Determination
One of the following:
•
Thermo Scientific™ Vanquish™ Variable
Wavelength detectors (VC-D40-A,
VF-D40-A)
•
Thermo Scientific™ Vanquish™ Diode Array detectors (VH
-D10-A, VF-D11-
A, VC-D11-A)
•
Thermo Scientific™ Vanquish™ Multiple Wavelength detector (VC
-D12-A)
Flow Cells Supporting
Automated Delay Volume
Determination
Flow cell, Standard, 11 µL, 10mm (P/N: 6077.0250)
Flow cell, Semi-micro, 2.5 µL, 7mm (P/N:6077.0360)
Flow cell, Standard, 13 µL, 10mm (P/N: 6083.0510)
Flow cell, Semi-Analyt., 5 µL, 7mm (P/N:6083.0520)
Flow cell, Semi-micro, 2.5 µL, 7mm (P/N: 6083.0530)
Flow cell, Semi-micro Bio, 2.5 µL, 7mm (P/N: 6083.0550)
LightPipe flow cell, Standard, 2 µL, 10 mm (P/N: 6083.0100B)
LightPipe flow cell, High Sensitivity, 13 µL, 60 mm (P/N: 6083.0200B)
Primary Wetted Material
Sample flow path: MP35N, PAEK, PAEK/PTFE composite, PEEK
Flush flow path: FEP, PEEK, ceramics
,
FFKM
Wash liquid flow path: Silicon, PP, FFPM, PEEK, PA
Waste flow: PP, PEEK, FEP, TPE
Biocompatibility
Yes
pH range: 2 - 12
Chloride concentration:
≤
1 mol/L
USB Communication
1 USB port (USB 2.0, "B" type connector)
1 USB hub with 3 ports (USB 2.0, "A" type connectors)
System Interlink
2 system interlink ports (RJ45-8 connectors)
Control
•
Basic control and functions are available for Chromeleon™ 7.2 SR5 and
Chromeleon™ 7.2.10
•
All functions software-
controlled, requires Thermo Scientific™
Chromeleon™ 7.3.1 CDS
•
Keypad: Mute Alarm, Light, Service, Valve, Wash, Rotate
•
Status Indicators: LED bar, Status LED
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...