8
•
Troubleshooting
Page 206
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Message and Code
Description and Remedial Action
Code 89. Leak sensor missing or
defective.
Here it means the front leak sensor.
Contact Technical Support for assistance.
To operate the device nevertheless, you can disable the leak sensor
functionality in the Chromeleon software by setting
Leak Sensor Mode
to
Disabled.
Code 9000. Firmware upgrading failed. Power cycle the device and retry.
If the issue persists, contact Technical Support.
Code 9001. Command rejected -
module is in error state.
The module cannot execute a command because it is in error state.
An internal error has occurred. Perform
Self Test
via FC ePanel: ePanel
FC tab
→
More Options
→
Service
→
SelfTest
button.
If the issue persists, contact Technical Support.
Code 9002. Rear leak sensor not
recognized or defective.
No rear liquid leak sensor is found or sensor malfunction.
Power cycle the device and retry.
If the error persists, contact Technical Support.
Code 9004. Liquid detected inside the
fraction collector compartment.
If liquid is detected inside the fraction collector compartment, the
warning message will be shown in the user-level audit trail, and drain
pump is triggered to drain out liquid to waste drain port. The device
keeps monitoring the status of the rear leak sensor and if the sensor
gets dry, the drain pump will stop working immediately.
Possible causes triggering the warning as follows:
•
1) Accumulated condensation if the device remains running at
cooling mode for a long time under a tough environment (e.g.
35°C with 80% RH).
•
2) Drain pump and its pump tubing broken, thereby failing to
drain out liquid inside the compartment.
•
2) Needle capillary broken.
•
3) Needle damaged or deformed.
•
4) Vial not present or vial broken.
•
5) Liquid spray out of needle wash port.
•
If the warning persists with drain pump working continuously,
contact Technical Support.
Code 9005. Carousel homing time out
Carousel
can’t
return to home position.
C
heck if the carousel is blocked by obstacles.
After resolving physical blockage, then perform self-test to recover the
issue.
If the issue persists, contact Thermo Fisher Scientific Technical Support
for assistance.
Code 9006. Needle horizon drive
homing time out
Needle drive
can’t return
to home position.
Check if the needle drive is physically blocked.
After resolving physical blockage, then perform
Self Test
to recover the
issue.
If the issue persists, contact Technical Support.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...