5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 85
5.8
Fraction Collection Valve
There are three capillaries or tubing that are required to assembly to the
fraction collection valve when installing the device for the first time.
•
Flush Solvent tubing (to flush solvent reservoir)
•
Delay capillary
•
Needle capillary (to needle unit inside fraction collector
compartment)
5.8.1
Port assignments of the Fraction Collection Valve
Figure 31: Fraction collection valve with connected capillaries except port
6
The fraction collection valve ports are assigned as follows:
Port
Connected Component
Port
Connected Component
1
to flush buffer loop
(pre-installed)
5
to Y-connector
(pre-installed)
2
to needle Capillary
6
Central port, inlet from detector
3
to Y-connector via air
bubble sensor (pre-installed)
7
to flush solvent
4
to flush buffer loop
(pre-installed)
8
to flush pump
(pre-installed)
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...