Follow these steps
1. Remove the front plate. See
Page 155).
2. Remove
needle wash outlet line
from the needle wash port inlet by
unscrewing the fitting and pull out tubing from the needle wash
pump outlet as shown below, then connect the
new needle wash
outlet line
to the free needle wash port inlet and the free needle
wash pump outlet and fit in the new needle wash outlet line into the
slot on the housing.
Figure 63: Needle wash outlet line
No.
Description
No.
Description
1
Needle wash port inlet
4
Needle wash port
2
The slot on the housing
5
Needle wash outlet line
3
Needle wash pump outlet
3. Re-install the front plate. See
(
Page 155).
1
2
3
4
5
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...