5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 83
No.
Description
A
Guide hole
1+2
Not to be used for wash liquid line; reserved for other tubing
3
Tubing guide for wash liquid line
4
Not to be used for wash liquid line; reserved for other tubing
Connecting the Needle Wash Inlet Line to the Needle Wash Reservoir
1. Connect the needle wash reservoir to the line:
a) Rinse the needle wash reservoir thoroughly with a high-purity
solvent.
b) Fill the needle wash reservoir with fresh needle wash liquid.
c) Feed the needle wash line through the retaining guide and
through an open hole in the cap of the needle wash reservoir.
The retaining guide keeps the tubing in place in the reservoir.
Close any open holes in the reservoir cap with cap plugs.
Figure 30: Securing the needle wash line with the reservoir cap
No.
Description
1
Needle wash inlet line
2
Retaining guide (
1/8’’
)
3
Reservoir cap
4
Cap plugs
2. Fill the needle wash reservoir with needle wash liquid. Mind the
requirements outlined in the previous section.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...