7
•
Maintenance and Service
Page 182
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Figure 82: Drain pump rotate counterclockwise to release the pump head
from pump motor
3. Unpack a replacement drain pump head with tubing.
4. Connect the condensation tubing in the device to the left tubing
connector on the replacement drain pump head with tubing.
5. Mount the replacement drain pump head and rotate it clockwise
slightly to the end by hand. When you hear a click, the replacement
drain pump head is successfully installed with pump motor together.
6. Insert the right-hand side waste tubing of the replacement drain
pump with tubing into the relevant hole and secure onto the front
leak tray.
7. In Chromeleon, perform the
DrainPumpHeadChanged
command.
8. Turn on the drain pump to check proper functionality.
NOTICE
Make sure to rotate the pump head gently and not damage the
pump tubing during the operation.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...