6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 129
How to Use the Tool
Figure 47: Vanquish Fraction Collector Delay Volume Calculator
1. Select the
Detector
configured in the Vanquish system from the
drop-down list.
2. Select the
Flow Cell Type
installed in the detector from the drop-
down list.
3. Select the installed
Delay Capillary IDxL
from the drop-down list.
This will automatically fill the
Delay Capillary Diameter
and
Delay
Capillary Length
.
If
Custom
is selected as
Delay Capillary IDxL
the values for
Delay
Capillary Diameter
and
Delay Capillary Length
need to be entered
manually.
4. Define the
Pump Flow Rate
to be applied in the instrument
method.
5. Define the
Detector Response Time
to be applied in the
instrument method.
6. The calculator displays the resulting
Delay Time
and
Delay
Volume
in the
Method Specific Delay
section.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...