6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 135
6.8.2
Adjust to InVial Pusher Position
Preparations
1. Open the left front door
.
2. Press the service button on keypad to move the needle unit into
service position.
NOTICE
To avoid personal injury, do not put hands into the
compartment when the needle unit is moving.
Follow these steps
1. Raise the lock pin upwards by hand.
T
he Lock tab will pop out from
the U-shape slot accordingly.
TIP
Do not press the lock pin. Just lift it up.
Figure 55:
Moving the lock pin up by hand
2. Release the lock pin. The pusher is released to
InVial
position.
3. Physically check that the needle tip does not protrude out of the
pusher and close the left front door.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...