7
•
Maintenance and Service
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 169
Figure 68: Installation of flush buffer loop and flush waste tubing
a) Unpack a replacement flush buffer loop and a replacement flush
waste tubing. Remove the cap from the Viper fittings.
b) Connect the Viper fittings to relevant ports of the fraction
collection valve and tighten them with the Viper knurled head.
c) Remove the knurls from the Viper fittings due to limited space at
the fraction collection valve.
10. Connect the capillary between flush pump and port 8 of fraction
collection valve:
Figure 69: Flush purge line connected to fraction collection valve and
flush pump