7
•
Maintenance and Service
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 173
3. Take one Viper knurl from other Viper capillaries as a tool for
following actions, e.g. from the delay capillary connecting detector
outlet and fraction collection valve central port. The knurl needs to
be fit with Viper fitting first before tightening or loosening it.
NOTICE
The device shall be powered off while replacing the needle
capillary.
Follow these steps
1. Remove the front plate. See
Page 155).
2. Loosen the attachment screw on needle capillary press plate with
the screwdriver till the needle capillary can be released from the
needle capillary press plate. The screw shall stay loose but not
removed from the device.
Figure 72: Loosening the attachment screw to release needle capillary
3. Loosen the needle capillary screw and disconnect the needle
capillary Viper fitting from the needle unit.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...