6
•
Operation
Page 140
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Situation after Shutdown
If no additive is used
If an addititive is used
Device shall be transported
or shipped after shutdown
Flush the system with
isopropanol.
Flush the system first with
several volumes (for
example, 1.0 mL/min for 10
minutes with the standard
system) of methanol and
water (50:50) to prevent
salt buildup in the fluidics. If
the solvents in the device
are not miscible with water,
use an appropriate
intermediate solvent.
Afterward, flush the system
with isopropanol.
1. Set the fraction collection valve to Waste position.
2. Turn off the pump flow to the device. Wait until the system pressure
is down to zero before you continue the shutdown of the device.
3.
If the device shall be transported or shipped after shutdown
Flush the needle wash system with an organic needle wash
solvent, such as isopropanol, before you interrupt operation to
avoid organic growth in the needle wash system.
Purge flush pump to flush buffer loop with an organic solvent,
such as isopropanol, before you interrupt operation to avoid
organic growth in the flush buffer loop.
Run an internal needle rinse cycle with an organic solvent, such
as isopropanol, before you interrupt operation to avoid organic
growth inside the needle capillary and needle of the fraction
collector.
4. Empty condensation drain system:
a) Leave the drain pump turned on until no condensation water is
present in the waste tubing of the drain pump. Then turn off the
drain pump.
b) Remove the sample racks and sample containers from the
fraction collector compartment.
c) If condensation or spilled samples are present in the fraction
collector compartment, clean and decontaminate the fraction
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...