7
•
Maintenance and Service
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 155
7.6
Removing the Front Plate
When
1. For needle capillary installation
2. For easier access to inner fluidic components
3. For easier access to needle unit
4. For replacement if case of broken front plate cable
Parts required in case repair
Replacement front plate
Tools required
Screwdriver, Torx T10
Preparations
Turn off the device with its main power switch.
NOTICE
The device shall be powered off while replacing the front plate.
Follow these steps
1. Loosen the two screws that attaches the removable front plate with
the screwdriver. The screws shall remain in the front plate.
Figure 57: Loosening the two screws in front plate
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...